Description
Advanced Conductivity Probe
Knowing the conductivity of liquid samples is a key parameter to control product quality throughout the production process. The optek ACF60 conductivity probe fulfils technical requirements of international regulations for reliable measurements ensuring quality and consistency.
Used For:
- Conductivity (DIN EN 27888 / ISO 7888 & ASTM D1125):
- 0 – 850 mS/cm
- User defined concentration measurement by sample conductivity
- Conductivity monitoring
- for fermenter / vessel control
- of CIP processes
- in TFF / Ultrafiltration
- of heat exchanger
- of pure, drinking and treatment water
- for interface detection in food and beverage
Typical Applications
Ultrafiltration
Chromatography
The conductivity probe ACS60 features a superior six-electrode, four-pole design. The arrangement of the four electrodes applying current around the two electrodes measuring the resulting voltage guarantees reliable and precise measurements in a broad conductivity range. In combination with the optek C8000 universal converter the ACS60 probe allows a wide dynamic range 0 – 850 mS/cm. This unique design provides a greatly reduced sensitivity to sensor fouling and polarization, which extends the probe’s lifetime.
ACS60 | ||
---|---|---|
Measurement | ||
Measuring Range | any measuring range between 0 – 10 μS/cm to 850 mS/cm | |
Protection | all parts have an IP rating of IP65 or higher | |
Process Adaption | ||
Material (Wetted) | PEEK (FDA, USP Class VI)
electrodes:
|
|
Port connection | OSP25-GS60 (similar Ingold-port)
nominal length:60 mm diameter:25 mm thread:G1-1/4 in. ISO 228/1 |
|
Port gasket | O-ring: EPDM (FDA / USP Class VI), … others on request | |
Process pressure | 20 bar (290 psi) up to + 50 °C (122 °F)
10 bar (145 psi) up to +100 °C (212 °F) 4 bar ( 58 psi) up to +135 °C (275 °F) |
|
Installation accessories |
|
|
Temperature Ratings | ||
Process temperature | permanent: -10 – 90 °C (14 – 194 °F)
peak 30 min/day: -10 – 135 °C (14 – 275 °F) |
|
Ambient temperature | operation:-10 – 40 °C (14 – 104 °F)
transport:-20 – 70 °C ( -4 – 158 °F) |
|
Temperature sensor | integrated Pt1000 RTD (IEC class A)
accuracy: ± 0.25 °C at 25 °C (77 °F) |
|
Accuracy | ||
0 – 10 μS/cm | calibrated: ± 1 % of measuring value ± 0.2 μS/cm
repeatability: ± 0.5% |
|
0 – 250 mS/cm | calibrated: ± 1 % of measuring value ± 0.2 μS/cm
repeatability: ± 0.5% |
|
250 – 500 mS/cm | calibrated: ± 2 % of measuring value ± 0.2 μS/cm
repeatability: ± 1% |
|
500 – 850 mS/cm | calibrated: ± 5 % of measuring value ± 0.2 μS/cm
repeatability: ± 3% |
|
Pressure and temperature ratings specified herein may be subject to limitations – see instruction manual. The appropriate choice of material for all wetted parts is the sole responsibility of the user. Data given are subject to changes without prior notice. |